Titan Krios 1 & 3:
Please find below the calibrated magnification table available for the our Krios 1 and Krios 3.
| Magnification | 350 kx | 270 kx | 215 kx | 130 kx | 105 kx | 81 kx | 64 kx |
| Krios 1 | 0.354 Å | 0.452 Å | 0.57 Å | 0.726 Å | 0.926 Å | ||
| Krios 3 | 0.355 Å | 0.456 Å | 0.574 Å | 0.732 Å | 0.934 Å | 1.192 Å | 1.515 Å |
Other pixel size available for the our Krios 1 and 3 – Not checked & Cryo-ET usage.
Titan Krios 2:
Please find below the calibrated magnification table available for the our Krios 2.
| Magnification | 250 kx | 195 kx | 155 kx | 120 kx | 96 kx | 75 kx |
| Krios 2 | 0.2925 Å | 0.3959 Å | 0.5082 Å | 0.658 Å | 0.83 Å | 1.063 Å |
Other pixel size available for the our Krios 2 – Not checked & Cryo-ET usage.
Glacios:
Please find below the calibrated magnification table available for the our Glacios.
| Magnification | 190 kx | 150 kx | 120 kx | 92 kx |
| Pixel Size | 0.72 Å | 0.94 Å | 1.2 Å | 1.5 Å |
| C2 Aperture | 20 µm | 20 / 30 µm | 30 µm | 30 / 70 µm |
| Beam Size (⌀) | 660 nm | 660 / 1.0 µm | 1.0 µm | 1.09 / 2.34 µm |
The microscope has currently four C2 apertures size : 20, 30, 70 and 150 µm. Any of those apertures selected for imagine should be centered. Currently the facility is using for SPA with EPU the preset with 20 or 30 µm for Grid Square, Hole/EucentricHeight and Data Acquisition, while the 150 µm aperture is usually reserved for atlas acquisition.
Reminder: Beam size on the 200 KeV TFS microscope is controled by the aperture size !
Table for Parallele illumination in Nanoprobe Gun lens 3 (last calibration 2026):
| Spot Size | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 |
| C2 Lens Current (%) | NA | 45.26 | 42.59 | 41 | 40.20 | 39.26 | 38.12 | 37.10 | 37.10 | 3.99 | NA |
Objective Aperture Resolution Cut-off:
| Objective | 100 | 70 | 50 | 30 |
| Glacios | 1.71 Å | 2.44 Å | 3.41 Å | 5.5 Å |
| Titan | 1.34 Å | 1.92 Å | 2.68 Å | 4.5 Å |